Field of Illumination and FOV - A technical overview...

Field of Illumination and FOV

A technical overview...

Here we define Field of View (FOV) by detector size and microscope objective, and Field of Illumination (FOI) relative to the detector and in the image plane.

In a microscope system, the camera is coupled via a C-mount adapter and located in a primary image plane (PIP). The PIP is our reference for the definition of FOV and FOI.

In the PIP, FOV is defined as the extent of the image sensor in X and Y dimensions. For consistency we define FOI as the extent of illumination in the PIP. This is convenient because we can easily calculate overlap between the two.

As mentioned above, it is common to match FOV and FOI, but with active illumination other factors such as Power Density (PD) or Resolution may also be important considerations. Mosaic is equipped a 2X zoom laser collimator so you can trade FOI for PD.

In a fluorescence microscope resolution is dominated by the objective lens which both illuminates and images the specimen. The objective numerical aperture (NA) and the wavelength of detected light (?) define Resolution, RXY by the Raleigh criterion as follows:

RXY = 0.61 * ?/NA;

In the table below RXY is computed for a range of objective lenses with ? = 500 nm.

Magnification (immersion medium) Numerical Aperture RXY (µm)
20 (air) 0.6 0.51
40 (air) 0.8 0.38
40 (WI) 1.0 0.30
60 (WI) 1.1 0.28
60 (Oil) 1.4 0.22
100 (Oil) 1.4 0.22


1/RXY is a good approximation of the maximum spatial frequency in the image. To capture all information in the image (e.g. with a CCD detector) we must sample at frequency F to avoid “aliasing errors”. This is known as the Nyquist criterion:

Sampling frequency, F = 2/(RXY)

Translating into the PIP with objective magnification of MO, we can compute the sensor pixel size required to fulfill the Nyquist criterion:

2 * Px = MO * RXY

Subject to application requirements, Nyquist may or may not be necessary, Using an objective lens of 100X, 1.4 NA we see that Neo, Clara, iXon3 885 and Luca R are all capable of achieving the Nyquist criterion: 2 * Px = 22 µm. While at 60X 1.4 NA, only Neo and Clara can provide small enough pixels.

MicroPoint and FRAPPA use Gaussian laser beams and Gaussian beams remain Gaussian with a theoretical minimum focus spot diameter of ?, where ? is the wavelength. This can only be achieved if the collimated laser beam fills the objective aperture.

The following variation on the Raleigh criterion provides a definition for spatial resolution SXY of the illumination system:

SXY = 1.22 ?f/D; f = focal length of objective,

D is beam diameter.

This is the size of the smallest object the microscope can resolve, sometimes called the diffraction limit, and is also the diameter of the smallest spot to which a collimated beam of incoherent light can be focused. The shape of the spot is an Airy disk or optical point spread function, PSF, characteristic of the system.

Magnification (immersion medium) Numerical Aperture SXY (µm)
20 (air) 0.6 1.02
40 (air) 0.8 0.76
40 (WI) 1.0 0.6
60 (WI) 1.1 0.56
60 (Oil) 1.4 0.44
100 (Oil) 1.4 0.40


The table above shows the approximate theoretical smallest spot sizes for a perfect microscope.

Power density, PD in the specimen plane is estimated from the ratio of beam power and area. Spectral transmission and chromatic errors in the microscope objective are critical to performance. To estimate specimen plane PDS, the Mosaic output beam power density is multiplied by the square of the magnification and the systems spectral transmission, T(?).

PDS = PDD* MO2 * T (?)

Channelrhodopsin2 (ChRh2) is a light activated cation channel which can be expressed in neurons and used to control behavior in host organisms, including mice, c. elegans and drosophila. Stimulation with blue light (~470 nm), the power density, for photo-activation of ChRh2 is in the range 0.1-10 mW mm-2 and has a wide dynamic range.

Multimedia Library
Application Images (102)
Show more
Application Movies (59)
Publications Database
The yeast Ste2p G protein-coupled receptor dimerizes on the cell plasma membrane
Amphiphilic Ferrocene-Containing PEG Block Copolymers as Micellar Nanocarriers and Smart Surfactants
Poly(glycerol sebacate) nanoparticles for encapsulation of hydrophobic anti-cancer drugs
Contrast Induced by a Static Magnetic Field for Improved Detection in Nanodiamond Fluorescence Microscopy
Heterogeneous expression of extracellular matrix molecules in the red nucleus of the rat
Flow of wormlike micellar solutions around confined microfluidic cylinders
Lateral Microscope Enables the Direct Observation of Cellular Interfaces and Quantification of Changes in Cell Morphology during Adhesion
Expansion and concatenation of nonmuscle myosin IIA filaments drive cellular contractile system formation during interphase and mitosis
Lung Microtissue Array to Screen the Fibrogenic Potential of Carbon Nanotubes
β-glucan microparticles targeted to epithelial APN as oral antigen delivery system
A Generic Polymer–Protein Ligation Strategy for Vaccine Delivery
Dyes as bifunctional markers of DNA hybridization on surfaces and mutation detection
Submicron thermal imaging of a nucleate boiling process using fluorescence microscopy
Evaluation of genetic damage in tobacco and arsenic exposed population of Southern Assam, India using buccal cytome assay and comet assay
Transient submicron temperature imaging based on the fluorescence emission in an Er/Yb co-doped glass–ceramic
Chloroquine-induced glioma cells death is associated with mitochondrial membrane potential loss, but not oxidative stress
Optical detection of two-color-fluorophore barcode for nanopore DNA sensing
Imaging potassium-flux through individual electropores in droplet interface bilayers
Anisotropic stress orients remodelling of mammalian limb bud ectoderm
An in vitro model of tissue boundary formation for dissecting the contribution of different boundary forming mechanisms