The iXonEM+ DU-860 back-illuminated EMCCD has single photon detection capability without an image intensifier, combined with greater than 90% QE of a back-illuminated sensor. Containing a 128 x 128 Frame Transfer CCD sensor from E2V Technologies, it enables charge to be multiplied on the sensor before it is read out, while utilizing the full QE performance of the CCD sensor. The absolute EMCCD gain of the camera can be varied linearly from unity up to a thousand times directly through the software, via a true quantitative EM gain scale. Andor are the first to offer this technology as a complete, software controlled system for scientific digital imaging applications. The system offers up to 10 MHz pixel readout rate, both EMCCD and conventional amplifier outputs and benefits from minimized dark current with unequalled thermoelectric cooling down to -100°C.
128 x 128
The smallest elements of a CCD detector grid.
24
3.1 x 3.1
Typical: 160000 Maximum: 220000
The well depth of a single pixel on the ccd sensor
800000
10MHz
515
Frame rate, or frame frequency, is the measurement of how quickly an imaging device produces unique consecutive images called frames
< 1 to 48 @ 10 MHz
Even single photon signals are amplified above the noise floor. Full QE of CCD chip is harnessed (no intensifier).
Absolute EMCCD gain selectable directly from a linear and quantitative scale.
Unique innovation that empowers the EMCCD to operate with market-leading acquisition efficiency through live cell microscopy software.
Critical for elimination of darkcurrent detection limit.
High detector sensitivity.
Quantitative accuracy at all speeds. Slower readout rate for enhanced 16-bit dynamic range.
Ideal for highly dynamic, low light experiments.
Critical for sustained vacuum integrity and to maintain unequalled cooling and QE performance, year after year.
Extended sensor dynamic range (readout speed dependent) and matched digitization (16-bit available) for quantization of dim and bright signals.
Giving lowest “spurious noise” specification; essential for single photon counting.
Essential for quantitative accuracy of dynamic measurements.
Easy means to record control dark images - excellent for optimization of experimental set-up.
Specialized readout mode for achieving sub-millisecond temporal resolution from EMCCD and CCD cameras. Intuitive Cropped Sensor definition for continuous rapid image or spectral acquisition.
BVBack Illuminated - Visible Anti Reflection Coating - 550nmAdd to wishlist
UVBBack Illuminated with UV Phosphor Add to wishlist
You will require one of these options to operate this camera
CCI-23PCI controller cardAdd to wishlist
3rd Party Software CompatibilityDrivers are available so the camera can operate through popular third party imaging software packages. Please contact us for a list of to see if you software is compatible Add to wishlist
iQThe best live cell imaging software just got betterAdd to wishlist
SDKAndor Software Development Kit (SDK)Add to wishlist
Solis (I)Solis-I for Imaging CamerasAdd to wishlist
XW-CHIL-150Ultra compact chiller unit for ultimate cooling performanceAdd to wishlist
XW-RECRRe-circulator for enhanced cooling performanceAdd to wishlist